STAT
From Agilent Multimeter Wiki
STAT?
Returns overall meter status
Effect
none
Response
"000001I00210L00104001"
- 1st digit: Average function 0=OFF 1=ON
- 2nd digit: Null function 0=OFF 1=ON
- 3rd digit: dB Function 0=OFF V=dBV M=dBM
- 4th digit: Change terminal 0=normal 1=alert
- 5th digit: Peak hold function 0=OFF 1=ON (PEAK button)
- 6th digit: Current percent 0=0-20mA 1=4-20mA
- 7th digit: "I" UNKNOWN
- 8th digit: Refresh hold function 0=OFF 1=ON (HOLD button)
- 9th digit: Zero temperature compensation 0=OFF 1=ON
- 10th digit: Beep selection 0=OFF C=300Hz F=600Hz 1=1200Hz 2=2400Hz
- 11th digit: Auto power off 0=OFF 1=ON
- 12th digit: Backlight 0=OFF 1=ON
- 13th digit: "L" UNKNOWN
- 14th digit: "0" UNKNOWN
- 15th digit: Amps terminal plug 0=Disconnected 1=Connected
- 16th digit: Rotary position (0-8 for U1252A, 0-7 for U1251A)
- 0=ACV, 1=V, 2=mV, 3=Res, 4=Diode, 5=Cap, 6=uA, 7= mA/A, 8=Square Wave (only U1252A)
- 17th digit: Output status 0=standby 1=operation (only U1252A)
- 18th digit: "4" UNKNOWN
- 19th digit: "0" UNKNOWN
- 20th digit: "0" UNKNOWN
- 21th digit: AutoRange 0=OFF 1=ON
- quotes included
Candidates for the UNKNOWN digits: battery level, fc derived path, input warning, meter mode, rate count, trigger source (the ones not identified from agilent gui properties tab).